November “In All Other News”: UK Update on Indefinite CE Marking, MIR Form Delayed, and more
3 December 2025
Manufacturer Incident Report (MIR) Form 7.3.1 Updated: Mandatory May 1, 2026
29 December 2025- Device (re)Classification
- Notified Body CE Marking Process
- Post-market Surveillance & Vigilance
- Clinical Evaluation Process
- Manufacturer Obligations (Article 10)
- Notified Body (NB) Fees
- Labeling (“Information Supplied with the Device”)
- Definition Changes (Article 2)
- International Reliance & Cooperation
- Other Changes That May Be of Interest
- What’s Next?
The EU Commission (EC) published its proposal to amend the MDR and IVDR: HERE
Commission Proposal COM(2025) 1023 includes significant revisions, in line with the EC’s public commitment to “simplify and streamline the regulatory framework”, “make it more cost-efficient and proportionate”, and “reduce the administrative burden”.
Device (re)Classification
The major changes proposed include:
- Software devices: Now starts as a Class I device and is up-classed based on the intended use. This is a major switch from current wording, which starts as Class IIa and then up-or-down classified based on intended use. Note that the MDCG guidance on classification certainly needs to be updated, to define what constitutes a “critical”, “serious”, or “non-serious” situation.
- Reusable Surgical Instruments: Now always Class I. This is the case even if it is intended for short-term use under Rule 7, and irrespective of which part of the body it contacts.
- Accessories for Active Implantable Devices: No longer automatically Class III. Instead, they are now classified in their own right.
- Well-Established Technologies: The newly defined term is included under Rule 8 (long-term implantable devices), exempting them from automatic Class III classification.
- Annex XVI Non-Medical Devices: Adds these into the classification scopes, e.g., Rule 9 for active devices.
Notified Body CE Marking Process
| Requirement | Current Obligation | Proposed Revision |
|---|---|---|
| CE Certificate Validity | Up to 5-years validity. | No expiration* (with periodic reviews**). |
| NB Surveillance Audits | At least per year. Must be onsite. | Can be every two years, where justified***. May conduct remote audits. |
| TD Surveillance: Class IIa/IIb Devices Class B/C IVD Devices | Technical documentation review required for representative devices. | Technical documentation reviewed only when concerns arise (risk-based). |
| Unannounced Audits | At least once every 5 years. | For cause. |
| Class III WET Devices | NB completes a technical documentation assessment for each device. | NB only reviews technical documentation for one representative device per generic device group†. |
| Class D IVD Devices | NB completes a technical documentation assessment for each device. | NB reviews technical documentation for one representative device per generic device group. |
| Class C IVD Devices | NB completes a technical documentation assessment for each device per generic device group. | NB reviews technical documentation for one representative device per generic device group. |
| Class IIb Non-Implantable WET Devices | NB completes a technical documentation assessment for each device. | NB reviews technical documentation for one representative device per category. |
| Class IIb Implantable WET Devices | NB completes a technical documentation assessment for each device. | NB reviews technical documentation for one representative device per generic device group. |
| Class IIa Devices Class B IVD Devices | NB review of at least one representative device for each category required. | NB review required of only one representative device for each category. |
| Class I Reusable Surgical Instruments | NB Certification required, covering at least the reusable aspects of the device. | NB Certification not required, if harmonised standards or Common Specifications covering all relevant aspects of the device are applied. |
| Class A Sterile IVD Devices | NB Certification required, covering at least the sterilization aspects of the device. | NB Certification no longer required. |
| Near-Patient IVDs | Annex IX Section 5.1 technical documentation & other usability/suitability requirements. | Normal conformity assessment process for their classification. |
*The validity of certificates shall not be limited in time, unless in exceptional cases where the notified body considers it necessary to limit the period of validity based on duly justified grounds. In those cases, the notified body shall indicate the period of validity on the certificate.
**During the validity of the certificate, the notified body shall carry out appropriate surveillance activities, including periodic reviews taking into consideration developments of the state of the art. Those reviews shall be proportionate to the risk class of the device.
*** The notified body shall carry out the surveillance audits and assessments once every 12 months. However, where justified in light of the results of previous surveillance audits and assessments, and in the absence of any concerns resulting from data from post-market surveillance or vigilance, the notified body shall carry out the surveillance audits and assessments only once every 24 months.
†Definitions per MDCG 2019-13, Rev.1:
- Category of devices: The relevant MDA/MDN and IVR/IVS codes codes according to Regulation (EU) 2017/2185 on the codes for the designation of notified bodies. There are 71 codes for MDR devices and 79 codes for IVD devices.
- Generic device group: The 4th level of the European Nomenclature on Medical Devices (EMDN) – combination of one letter plus 6 digits (MDR). The 3rd level of the EMDN – combination of one letter plus 4 digits + IVP code (IVDR).
Post-market Surveillance & Vigilance
Post-Market Surveillance
The Periodic Safety Update Reporting (PSUR) timeframe has been reduced.
| Device Classification | Current Requirement | Proposed Revision |
|---|---|---|
| Class IIb/III Devices Class C/D IVD Devices | PSUR updated at least once (1) per year. | PSUR updated in the first year after the CE Certificate issuance and every two (2) years thereafter. |
| Class III & IIb Implantable Devices (except WET) | NB reviews as each PSUR is generated. | NB reviews during surveillance audits. |
| Class III & IIb Implantable WET Devices | NB reviews PSUR. | NB does not review PSUR. |
| Class IIa Devices | Update at least every two (2) years. | Update “when necessary”. |
| Class D IVD Devices | NB reviews as each PSUR is generated. | NB reviews during surveillance audits. |
| Class C IVDs | PSUR made available to NB upon request. | NB no longer needs access to PSUR. |
Vigilance
The reporting timeframe was reduced for “all other” serious incidents. It is reduced back to the MDD-IVDD era reporting requirements.
| Type of Serious Incident | Current Requirement | Proposed Revision |
|---|---|---|
| Serious public health threat | No later than 2 calendar days after the manufacturer becomes aware | None |
| Death or unanticipated serious deterioration in state of health | No later than 10 calendar days after the manufacturer becomes aware | None |
| All other | No later than 15 calendar days after the manufacturer became aware | No later than 30 calendar days after the manufacturer became aware |
Clinical Evaluation Process
Article 61 expanded to:
- Allow additional non-clinical sources of data, in particular “in vitro, ex vivo, in silico testing, computational modeling or simulation”. In other words, where duly justified, manufacturers may rely on additional sources of non-clinical testing without requiring, for example, data from a clinical investigation.
- Manufacturers may now consider data from other devices within their generic device group when justifying reliance on non-clinical testing methods. This can reduce the need for redundant testing when launching variants or line extensions of existing products.
- Remove Class IIa implantable devices from Clause 4’s clinical investigation requirement.
- Remove the specific list of WET devices and replace it more generally with ‘well-established technology devices’, which is newly defined in Article 2.
- Allow manufacturers to include PMCF findings directly in the updated Clinical Evaluation Report, without needing to generate a separate PMCF evaluation report.
Manufacturer Obligations (Article 10)
Below are some of the changes proposed:
- Quality Management System (QMS): Reduced the language around QMS requirements, deleting the entire section outlining the minimum requirements (bullets (a)-(m)). Manufacturers must still have an appropriate QMS covering the “structure, responsibilities, procedures, processes and management resources required to implement the principles and actions necessary to achieve compliance with the provisions of this Regulation.”
- Language for Labeling: New clause added. While it is under the Manufacturer’s Obligations, it is directed to national competent authorities, asking them to “consider accepting another official language of the Union in which the information is made available, taking into consideration the technical knowledge, experience, education or training of the average intended user(s)”.
- “Virtual” Manufacturers: Adds a statement for manufacturers who have their devices designed or manufactured by someone else. Requires that the manufacturer ensure that the technical documentation is available from their supplier(s), can be made available to the Competent Authorities upon request, and that the manufacturer shall draw up (and keep current) the technical documentation, in particular, those referred to in Section 2 of Annex II and Annex III.
- Financial Coverage for Liability Claims: This clause has been deleted. This clause currently legally obligates the manufacturer to have sufficient financial coverage for liability claims. While deleted in the reform package, manufacturers are still subject to the EU Product Liability Directive 85/374/EEC.
Notified Body (NB) Fees
The MDR/IVDR already required NBs to publicly publish their list of standard fees. However, the following additional changes are proposed:
- The Commission will create a dedicated, public webpage, making reference to each NBs fees.
- A fee reduction of at least 50% shall be provided for micro-sized manufacturers and for orphan devices.
- A fee reduction of at least 25% shall be provided for small enterprises.
- Micro-and-small enterprises shall have the option to defer payment until the conformity assessment process has been finalized.
- The Commission may adopt implementing acts to specify the structure and level of NB fees.
- NBs shall ensure that micro, small, and mid-sized manufacturers are treated equally to larger organizations.
- NBs shall reply to a manufacturer’s conformity assessment request within 15-days of receipt of the request.
- The authorities responsible for NBs may, in the interest of public health or patient health or safety, instruct the NB to accept a manufacturer’s conformity assessment request (as long as it falls within the NB’s scope of designation).
The definitions for micro, small, and medium sized enterprises are defined in Recommendation 2003/361/EC.
| Micro Enterprise | Small Enterprise | Medium Enterprise |
|---|---|---|
| Employs fewer than 10 people and its annual turnover and/or annual balance sheet total does not exceed EUR 2 million. | Employs fewer than 50 people and its annual turnover and/or annual balance sheet total does not exceed EUR 10 million. | Employs fewer than 250 people and its annual turnover does not exceed EUR 50 million, and/or its annual balance sheet total does not exceed EUR 43 million. |
Labeling (“Information Supplied with the Device”)
Some changes were proposed, including:
- MDR Annex I, Section 23.1/IVDR Annex I, Section 20.1 REVISED: Deletes the requirement that the Instructions for Use “shall, if the manufacturer has a website, be made available and kept up to date on the website”.
- IVDR Annex I, Section 20.1(f) REVISED: Removes the eIFU exception for near-patient testing.
- MDR Annex I, Section 23.2(0) DELETED: “If the device is a single-use device that has been reprocessed, an indication of that fact, the number of reprocessing cycles already performed, and any limitation as regards the number of reprocessing cycles”
- MDR Annex I, Section 23.2(q) REVISED: Adds “or an accessory for a medical device”. Currently, the EU harmonized “MD” symbol can be used to note a product is a medical device. Presumably the Harmonized Standard EN ISO 15223-1:2021 must be updated with a new symbol for accessories, or to note it is also applicable to accessories for medical devices.
- IVDR Annex I, Section 20.2(e) REVISED: Adds ” or an accessory for an in vitro diagnostic medical device”. Currently, the EU harmonized “IVD” symbol can be used to note a product is a medical device. Presumably the Harmonized Standard EN ISO 15223-1:2021 must be updated with a new symbol for accessories, or to note it is also applicable to accessories for medical devices.
Definition Changes (Article 2)
| Newly Added | Modified | Deleted |
|---|---|---|
| -Well-established technology device -Combined study -Regulatory sandbox -Union regulatory sandbox | -Medical device (minor) -Generic device group (moderate) -Clinical performance (significant) | -Nanomaterial Particle -Agglomerate -Aggregate |
International Reliance & Cooperation
The Commission added articles on International Reliance and Cooperation mechanisms.
This is a significant addition to the Regulations, as the newly proposed article allows the Commission to “participate in bilateral or multilateral reliance mechanisms or reliance programmes in the field of medical devices and in vitro diagnostic medical devices”.
This would be in line with the international approach taken by other regulators, such as the UK MHRA’s upcoming International Reliance regulation, and Switzerland considering acceptance of FDA devices (both requiring abridged review by a third-party certifying body).
Other Changes That May Be of Interest
- Distance Sales: Article 6 updated so that manufacturers must provide the labeling information outlined under the MDR Annex I, Section 23.2 “Information on the label”, points (a)-(d) and (m) / IVDR Annex I, Section 20.2 “Information on the label”, points (a)-(e) and (m) and access to the Instructions for Use.
- Authorized Representative Obligations: Article 11 was modified to delete the following: 1) clause which prevents manufacturers from delegating certain obligations to the Authorized Representative, such as UDI and device registration requirements, and 2) clause which makes the Authorized Representative jointly and legally liable for the defective devices, if the manufacturer defaults on their obligations.
- Person Responsible for Regulatory Compliance (PRRC): Article 15 amended to remove the minimum requirements to act as a PRRC. It now simply states that the PRRC must possess the requisite expertise in the field of medical devices. Further, micro and small enterprises who outsource the PRRC role, are no longer required to have them “permanently and continuously” at their disposal.
- Single-use Devices and Reprocessing: Article 17 updated to note that manufacturers must provide a justification rationale as to why the device cannot be safely reused. And where the product can be reused, the manufacturer must provide information about the reprocessing process in the Instructions for Use. The article still allows refurbishing of single-use devices, but greatly reduces the language around this allowance.
- Implant Cards: Article 18 modified to state the implant card may be provided in electronic or digital format. Further, that implants which are ‘well-established technology devices’ are exempt from this requirement.
- Declarations of Conformity (DoC): Article 19 edited to allow electronic provision of DoCs.
- IVD Kits: New Article 19a in the IVDR. Content aligned to MDCG 2024-11, but would now be legally binding.
- Systems and Procedure Packs: Article 22 amended to note the Statement must include identification of 1) the devices, 2) where appliable, other products included in the system or procedure pack, 3) where applicable, the device Basic UDI-DI, and 4) where applicable, identification of the Notified Body involved in the sterilization activities, if the system/procedure pack producer sterilizes the system or procedure pack in order to place them on the market.
- Summary of Safety & Clinical Performance (SSCP): Article 32 amended. SSCPs no longer required for Class IIa implantable devices or ‘well-established technology devices’. Nor is separate validation required by the Notified Body.
- Release of Remaining EUDAMED Modules: Article 33 amended to give the Commission leeway to not release all six modules within EUDAMED. However, modules developed outside EUDAMED (e.g., Clinical Investigations) must be interoperable with it.
- Notified Body Disputes: Article 35 amended to provide a clearer mechanism (including 90-day decision timeframe) for disputes between a manufacturer and Notified Body.
- UDI Issuing Agencies: Should take micro, small, and mid-sized companies into consideration, including offering preferential conditions for its UDI assignment system.
- National Registration Databases: EUDAMED’s development shall consider compatibility with national databases. If a national distributor database requires information on devices, the database should allow for that data to be retrieved from EUDAMED.
- Structured Dialogues with Notified Bodies: Currently, these are extremely limited in scope, especially for pre-application discussions. However, the Commission proposes to expand pre-submission discussions, more like UK MHRA Regulatory Advice and US FDA Q-Sub meetings.
There are numerous other revisions not described in this article. For example, significant changes to the Notified Body designation process, new articles on breakthrough devices and orphan devices, regulatory sandboxes, and clinical investigations for combined studies. As well as many minor changes, for example, a clearer definition for Basic UDI-DI.
The above article highlights the items we consider as having the most direct impact, on the majority of medical device manufacturers.
What’s Next?
For an ‘Ordinary Legislative Procedure’, such as this one, the next step is European Parliament Review, assigned to the Committee on Environment, Public Health and Food Safety (ENVI). In parallel, the Council of the EU will also initiate its review through its Working Party on Pharmaceuticals and Medical Devices.
After these reviews, the European Parliament, Council of the EU, and Commission will enter into informal trilogue negotiations to reconcile differences and agree on a final compromise text.
General timing for an Ordinary Legislative Procedure can be 2+ years. However, this amendment could be fast-tracked, due to: 1) Parliament’s October 2024 resolution calling for urgent reforms, 2) the Commission’s stated concern about medical device availability, and 3) looming MDR (December 2027/2028) and IVDR (December 2028/2029) transition deadlines.
Last, it is important to note that the proposal will undergo review and amendment by Parliament and Council, with input from industry stakeholders, competent authorities, and patient organizations. The final adopted text may differ significantly from this December 2025 Commission proposal.