EU MDR/IVDR Vigilance Assessment Tool A step-by-step decision tool to help manufacturers assess if an incident involving their medical device requires reporting under EU MDR/IVDR requirements. Need Help?

JUMP TO VIGILANCE ASSESSMENT TOOL

Manufacturers are responsible for assessing incidents that occur with their products on the market and determining whether those incidents require reporting to a Competent Authority. Making this determination requires working through a structured series of questions: Was there an incident? Was it serious? Is the device a contributory cause? What are the timelines to report this incident type, if applicable?

Our step-by-step MDR/IVDR Vigilance Reporting Assessment Tool guides manufacturers through this decision-making process. It includes help text, such as embedded regulatory definitions and MDCG guidance, to support each determination.

What’s Included in the Vigilance Reporting Assessment Tool?

  • Guides users through the relevant questions required to determine whether a situation is reportable.
  • References the applicable MDR/IVDR articles and MDCG 2023-3 at each step.
  • Provides a clear action statement at the end of the assessment, i.e., whether reporting is required or not, and why.
  • Reminds manufacturers of their QMS documentation obligations regardless of whether the situation is reportable.
  • Flags trend reporting obligations under MDR Article 88/IVDR Article 83 where relevant.
  • Allows the user to have the final assessment emailed as a PDF for their records. The PDF output can include the following information: the date the incident occurred, device name, summary of the incident, full decision path taken to obtain the end result, action items as applicable (e.g., timeframe to report, reminder to document the assessment), and other general notes as needed.

How the Assessment Works

The tool walks through the following questions in sequence:

Step 1: Does the situation meet the definition of an incident?

The tool provides the full regulatory definition and additional context to assist in making this determination.

Step 2: Is the incident serious?

If the situation meets the definition of an incident, the tool assesses whether it meets the threshold for a serious incident per MDR Article 2(65) / IVDR Article 2(68). The tool references MDCG 2023-3 for additional context on what constitutes a serious deterioration in state of health, an unanticipated serious deterioration, and a serious public health threat.

Step 3: Is there a causal relationship with the device?

Per MDCG 2023-3, reporting is required where a causal relationship exists or cannot reasonably be excluded. The tool provides the relevant MDCG guidance on how to assess this, and notes that in case of doubt, the manufacturer must report.

Outcome: Action Required

At the conclusion of the assessment, the tool provides a clear statement of what action, if any, is required, including applicable reporting timelines and QMS documentation obligations.


Vigilance Contacts

For information on how to report and the applicable Competent Authority contacts, visit our other vigilance resource.


Reporting Timelines

If a serious incident has been reported or is suspected and a causal relationship established or not excluded, the manufacturer must report within the following timeframes:

Type of Serious IncidentDeadline to Report
Serious public health threatNo later than 2 calendar days after the manufacturer becomes aware
The death of a patient, user, or other person; an unanticipated serious deterioration of a patient’s, user’s, or other person’s state of healthNo later than 10 calendar days after the manufacturer becomes aware
The temporary or permanent serious deterioration of a patient’s, user’s, or other person’s state of healthNo later than 15 calendar days after the manufacturer becomes aware

Vigilance Assessment Tool



A Note on This Tool

This vigilance tool provides a preliminary assessment to help manufacturers understand the reporting obligations under EU MDR 2017/745 and IVDR 2017/746. This tool offers a starting point for internal planning and should not be considered a binding regulatory determination or comprehensive interpretation of the regulation.

For a definitive confirmation as to whether an incident should be reported, manufacturers should review the applicable regulation further or seek professional regulatory advice. Casus Consulting has prepared this tool with care but does not guarantee absolute accuracy or assume legal liability for decisions made based on these results. For specific regulatory guidance tailored to your device, please contact Casus Consulting directly.